ichappsseo's Posts

Preparation of silver nitrate solution

Pour about 75 ml of Distilled water into a 250 ml capacity beaker. Weigh 5.0 grams of Silver Nitrate LR and transfer slowly, into above said beaker, through it‘s sidewalls, while gentle mixing by using a glass rod. Transfer prepared solution into a 100 ml volumetric flask, by using a glass funnel. Rinse the beaker and funnel with few ml of Distilled water, and allow to add into the contents of volumetric flask.

https://ichapps.com/qms/view/preparation-of-silver-nitrate-solution-5-percent

continue reading
1 visits |0 Comments

Operation Cleaning and Calibration of Double Door Autoclave

Check the water level before starting the autoclave. If the level is down, then add purified water upto the level mark. Open the door at non-sterile area and load the necessary items to be sterilized as per appendix – I / appendix –II as required. Place the autoclave indicator strips at different places in the autoclave for each cycle to indicate the proper sterilization by color differentiation of the strip. 

https://ichapps.com/qms/view/operation-cleaning-and-calibration-of-double-door-autoclave

continue reading
1 visits |0 Comments

preparation of sodium hydroxide solution 10.0%

Pour about 75 ml of distilled water into a 250 ml capacity beaker. Weigh 10.0 grams of Sodium Hydroxide Pellets LR and transfer slowly, into above said beaker, through it‘s sidewalls, while gentle mixing by using a glass rod. Transfer prepared solution into a 100 ml volumetric flask by using a glass funnel.

https://ichapps.com/qms/view/preparation-of-sodium-hydroxide-solution-100-percent

continue reading
3 visits |0 Comments

Operation Cleaning And Calibration Of Single Door Autoclave

Check the water level before starting the autoclave. If the level is down, then add purified water upto the level mark. Open the door and load the necessary items to be sterilized as required. Place the autoclave indicator strips at different places in the autoclave for each cycle to indicate the proper sterilization by color differentiation of the strip.

https://ichapps.com/qms/view/operation-cleaning-and-calibration-of-single-door-autoclave

continue reading
3 visits |0 Comments

preparation of sodium hydroxide solution dilute

Pour about 75 ml of distilled water into a 250 ml capacity beaker. Weigh 5.0 grams of Sodium Hydroxide Pellets LR and transfer slowly, into above said beaker, through it‘s sidewalls, while gentle mixing by using a glass rod. Transfer prepared solution into a 100 ml volumetric flask by using a glass funnel.

https://ichapps.com/qms/view/preparation-of-sodium-hydroxide-solution-dilute

continue reading
2 visits |0 Comments

Procedure for Operation Cleaning and Calibration of Hot Air Oven

Connect power plug to the socket. Put ON the Mains on the control panel, Orange lamp inside it glows and temperature controller as well as the PC / Interface displays and the inside Incubator motor ON. Put ON the heating switch on the control panel, Dry heater will ON and green indicator lamp glows. Connect power plug to the socket. Put ON the Mains on the control panel, Orange lamp inside it glows and temperature controller as well as the PC / Interface displays and the inside Incubator motor ON. Put ON the heating switch on the control panel, Dry heater will ON and green indicator lamp glows. 

https://ichapps.com/qms/view/procedure-for-operation-cleaning-and-calibration-of-hot-air-oven

continue reading
3 visits |0 Comments

Operation and Calibration of (VACUUM OVEN CINTEX CIC 61)

Basic Operation: Before using the Vacuum oven, check the validity of calibration. Switch on the instrument by the “MAINS” switch on the control panel of the instrument. The control panel consists of Microprocessor based Temperature controller, Three heat control switch (Rotatory switch) to control heating rate control, Fuse, Vacuum gauge.

https://ichapps.com/qms/view/operation-and-calibration-of-vacuum-oven-cintex-cic-61

continue reading
1 visits |0 Comments

Operation and Calibration of UV Cabinet Advance

Before using  ensure the validity of calibration. Switch  on  the  power. Switch on the normal light and open the front door pulling towords the front side to keep the TLC plates. Place the TLC plate in UV chamber horizontally on the base of the chamber. 

https://ichapps.com/qms/view/operation-and-calibration-of-uv-cabinet-advance

continue reading
1 visits |0 Comments

Procedure for Cleaning of Microbiology Lab

Prepare the disinfectant solution as per the current version of SOP No. xxxxx... Take the disinfectant solution in two buckets and mark the bucket as 1 and 2. Enter into the Laminar airflow room through change rooms as per the current version of SOP no.xxxxx...

https://ichapps.com/qms/view/procedure-for-cleaning-of-microbiology-lab

continue reading
1 visits |0 Comments

Procedure for Operation and Calibration of MAS 100 Air Sampler

Place the MAS –100 Air Sampler on a firm support. Press  “ YES” to turn ON the MAS –100. After power – on, the MAS –100 displays the software with the current battery charge, followed by the most recent aspirated volume. Check the instrument battery.

https://ichapps.com/qms/view/procedure-for-operation-and-calibration-of-mas-100-air-sampler

continue reading
1 visits |0 Comments
[]